Enumeration of Major Peripheral Blood Leukocyte Populations for Multicenter Clinical Trials Using a Whole Blood Phenotyping Assay
Published on: September 16, 2012
Comparison of Flow Cytometric Methods for the Enumeration of Residual Leucocytes in Leucoreduced Blood Products: A
Yang Zeng1, Michelle Dabay2, Virginia George2
1BD Life Sciences, Medical Affairs and Corporate Clinical Development, San Jose, California 95131.
Abstract:
The BD FACSVia™ System features novel designs in hardware, software, and instrument QC. We compared the performance of the BD FACSVia System using the BD Leucocount™ kit with the BD FACSCalibur™ flow cytometer. Leucoreduced platelet (PLT, n = 252) and red blood cell (RBC, n = 278) specimens were enrolled at four sites. Each specimen was stained in four tubes using the BD Leucocount kit reagents and acquired on the two systems. BD Leucocount Control cells (high and low) were used to evaluate the inter-site reproducibility on the BD FACSVia System at three sites over 20 days. Deming regression and Bland-Altman analysis were performed to determine the WBC absolute counts on the BD FACSVia System vs. the BD FACSCalibur system. Assay accuracy for the range of 0-350 WBCs/µl was adequate. For samples with <25 WBCs/µl, the bias with 95% limits of agreement was 0.136 (-1.897 to 2.169) WBC/µl for PLTs (n = 184) and 0.170 (-2.025 to 2.365) WBC/µl for RBCs (n = 193). For inter-site reproducibility, the CV% was 6.46% (upper 95% CI 7.16%) for the PLT high control and 9.49% (10.52%) for the PLT low control. The CV% was 7.51% (8.32%) for the RBC high control and 10.76% (11.92%) for the RBC low control. The BD FACSVia System reported equivalent results of WBC absolute counts for leucoreduced PLT and RBC samples compared to the BD FACSCalibur system. The inter-laboratory reproducibility of the BD FACSVia System met study specifications. © 2018 The Authors. Cytometry Part A Published by Wiley Periodicals, Inc. on behalf of ISAC.
More Related Videos
06:28A Simple Flow Cytometric Method to Measure Glucose Uptake and Glucose Transporter Expression for Monocyte Subpopulations in Whole Blood
Published on: August 12, 2016
13:20Flow Cytometric Measurement Of ROS Production In Macrophages In Response To FcγR Cross-linking
Published on: March 7, 2019
Related Concept Videos
Blood Flow
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
Residual Plots
When the residual values are plotted against the variable x, it is called a residual...
Residual Stresses
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Applications of Integration to Find Blood Flow