IFCC Working Group Recommendations for Assessing Commutability Part 3: Using the Calibration Effectiveness of a
Jeffrey R Budd1, Cas Weykamp2, Robert Rej3
1Beckman Coulter, Chaska, MN.
Abstract:
A process is described to assess the commutability of a reference material (RM) intended for use as a calibrator based on its ability to fulfill its intended use in a calibration traceability scheme to produce equivalent clinical sample (CS) results among different measurement procedures (MPs) for the same measurand. Three sources of systematic error are elucidated in the context of creating the calibration model for translating MP signals to measurand amounts: calibration fit, calibrator level trueness, and commutability. An example set of 40 CS results from 7 MPs is used to illustrate estimation of bias and variability for each MP. The candidate RM is then used to recalibrate each MP, and its effectiveness in reducing the systematic error among the MPs within an acceptable level of equivalence based on medical requirements confirms its commutability for those MPs. The RM is declared noncommutable for MPs for which, after recalibration, the CS results do not agree with those from other MPs. When a lack of agreement is found, other potential causes, including lack of calibration fit, should be investigated before concluding the RM is noncommutable. The RM is considered fit for purpose for those MPs where commutability is demonstrated.
Related Concept Videos
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Inertial Frames of Reference
Non-inertial Frames of Reference
Framing Effects