Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
[The Measurement Precision and Accuracy of T1 Mapping Using Polarity Corrected (PC) TI Prep Tool]
Hironobu Ishikawa1, Shinya Seino1, Hideaki Takasumi1
1Department of Radiology, Public University Corporation Fukushima Medical University.
Abstract:
The aim of this study was to evaluate the measurement precision and accuracy of T1 mapping using a polarity corrected (PC) TI prep tool, which was based on fast field echo (FFE) and obtained one data point with one inversion recovery (IR) pulse. A phantom was used consisting of eight materials with different Gd concentrations. T1 mappings were measured by changing the trigger interval and the inversion time (TI) interval. The T1 mapping measurement precision using the PC TI prep tool increased as the trigger interval was made longer. The measurement precision didn't depend on the interval of TI. On the other hand, when the trigger intervals are more than 1000 ms, the measurement accuracy was less than approximately 8%. By setting the optimal end of TI, the T1 mapping using a PC TI prep tool could measure the T1 value precisely and accurately.