Refining non-invasive techniques to measure intracranial pressure: comparing evoked and spontaneous tympanic membrane
Laurie C Finch1,2, Robert J Marchbanks2,3, Diederik Bulters1,4
1Faculty of Medicine, University of Southampton, University Hospital Southampton NHS Foundation Trust, Tremona Road, Southampton, SO16 6YD, United Kingdom.
Objective:
Tympanic membrane displacements (TMDs) are used to non-invasively gauge inner-ear fluid pressure. Inner-ear fluid pressure equalizes with intracranial pressure (ICP) via the cochlear aqueduct and therefore TMDs can indirectly evaluate ICP. We studied the relationship between two TMD modalities, evoked and spontaneous. Evoked TMD is a reflex response to an auditory stimulus and the established stapes-footplate mechanism explains how evoked TMDs change with ICP. Spontaneous TMD refers to a pulsatile TMD waveform expressed in the form of pulse amplitudes (TMD-PAs), the origins of which are poorly understood. We investigated whether both modalities respond similarly to an ICP change, suggesting a common mechanism.
Approach:
ICP was manipulated in 20 healthy volunteers by a postural change from sitting (lower ICP) to supine (higher ICP). Differences between paired sitting and supine TMD results generated ΔEvoked and ΔSpontaneous values.
Main Results:
Evoked TMDs became more inward on lying supine while spontaneous TMDs became more outward. There was no evidence of a correlation between ΔEvoked and ΔSpontaneous (Right ears: r = -0.38, p = 0.10, 95% CI -0.75 to 0.21; Left ears: r = 0.34, p = 0.16, 95% CI -0.17 to 0.75).
Significance:
This suggests the stapes-footplate mechanism is not the primary mechanism explaining how spontaneous TMDs respond to changing ICP.
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Spontaneity
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
Displacement Current