Injecting without pressing a button: An exploratory study of a shield-triggered injection mechanism
Eric Zijlstra1, Hans-Veit Coester1, Tim Heise1
1Profil, Neuss, Germany.
The new DV3316 shield-triggered pen-injector demonstrated high injection success and user confidence, comparable to FlexPen, with a majority preferring the DV3316 device for diabetes self-injections.
Area of Science:
- Medical Devices
- Diabetes Management
- Injection Technology
Background:
- Pen-injectors are crucial for subcutaneous drug delivery in diabetes management.
- User experience and injection success are key factors in device adoption.
- Evaluating novel injection mechanisms is essential for improving patient self-care.
Purpose of the Study:
- To assess the injection success rate of a novel shield-triggered pen-injector (DV3316).
- To evaluate user perception, including confidence, pain, and device experience, of the DV3316 compared to FlexPen.
- To determine overall participant preference between the two pen-injector devices.
Main Methods:
- An exploratory, randomized controlled trial involving 150 injection-experienced individuals with diabetes.
- Participants self-administered placebo injections using the DV3316 and FlexPen devices.
- Injection success, participant confidence, pain, and device experience were systematically evaluated.
Main Results:
- High injection success rates were observed for both DV3316 (97.0%) and FlexPen (99.7%).
- Participant confidence and median injection pain scores were comparable between the two devices.
- A significant majority (55%) preferred the DV3316 over the FlexPen (21%).
Conclusions:
- The shield-triggered DV3316 pen-injector achieves high injection success in experienced individuals.
- Patient confidence and injection pain associated with DV3316 are similar to the established FlexPen.
- The DV3316 shows potential for improved user preference in diabetes self-injection.
Related Concept Videos
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Nonlinear Pharmacokinetics: Drug Elimination for IV Bolus Injection
Following the administration of a single intravenous (IV) bolus injection, we can determine the concentration of the drug in the plasma at any given time. This calculation is achieved using a specific equation that integrates the values of Vmax and KM.
We can also...
Drug Accumulation During Multiple Dosing: Repetitive IV Injections
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Reaction Mechanisms
For instance, the decomposition of ozone appears to follow a mechanism with two steps: