Related Experiment Video
Updated: Feb 15, 2026

Author Spotlight: Utilizing Venoplasty Balloon Model in Rodents to Simulate Surgical Interventions for Deep Veins
Published on: May 24, 2024
Differences in radial expansion force among inferior vena cava filter models support documented perforation rates
J Eli Robins1, Ihab Ragai2, Dean J Yamaguchi1
1Division of Vascular Surgery, Department of Cardiovascular Sciences, East Carolina University Brody School of Medicine, Greenville, NC.
Objective:
Inferior vena cava (IVC) filters are used in patients at risk for pulmonary embolism who cannot be anticoagulated. Unfortunately, these filters are not without risk, and complications include perforation, migration, and filter fracture. The most prevalent complication is filter perforation of the IVC, with incidence varying among filter models. To our knowledge, the mechanical properties of IVC filters have not been evaluated and are not readily available through the manufacturer. This study sought to determine whether differences in mechanical properties are similar to differences in documented perforation rates.
Methods:
The radial expansion forces of Greenfield (Boston Scientific, Marlborough, Mass), Cook Celect (Cook Medical, Bloomington, Ind), and Cook Platinum filters were analyzed with three replicates per group. The intrinsic force exerted by the filter on the measuring device was collected in real time during controlled expansion. Replicates were averaged and significance was determined by calculating analysis of covariance using SAS software (SAS Institute, Cary, NC).
Results:
Each filter model generated a significantly different radial expansion force (P < .001), and force was distributed at significantly different rates (P < .001) during expansion. The largest radial expansion force at minimal caval diameter was seen in the Cook Platinum filter, followed by the Cook Celect and Greenfield filters. Radial force dispersion during expansion was greatest in the Cook Celect, followed by the Cook Platinum and Greenfield filters.
Conclusions:
Differences in radial expansion forces among IVC filter models are consistent with documented perforation rates. Cook Celect IVC filters have a higher incidence of perforation compared with Greenfield filters when they are left in place for >90 days. Evaluation of Cook Celect filters yielded a significantly higher radial expansion force at minimum caval diameter, with greater force dispersion during expansion.
Related Concept Videos
Methods of Documentation VI: Case Management Model
For example, a patient with a chronic...
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Active Filters
Radial System Protection
In a radial system with a fault downstream of the third breaker, ideally, only the third breaker will open, isolating the fault and interrupting the load connected beyond it. The second breaker has a longer delay setting,...
Electromotive Force
Esophageal Perforation-I: Introduction
The location of esophageal perforation can vary, occurring anywhere along the esophagus....