Determination of phthalate esters in physiological saline solution by monolithic silica spin column extraction method
Lu Lu1,2, Yuki Hashi3, Zhi-Hua Wang1
1State Key Laboratory of Chemical Resource Engineering, Beijing University of Chemical Technology, Beijing 100029, China.
A new eco-friendly method using monolithic silica spin column extraction (MonoSpin-SPE) and UFLC-MS accurately quantifies six phthalate esters in saline. Dimethyl- and diethyl-phthalate were detected in commercial saline samples.
Area of Science:
- Analytical Chemistry
- Environmental Science
Background:
- Phthalate esters are common plasticizers with potential health concerns.
- Accurate quantification of phthalates in medical solutions like physiological saline is crucial.
Purpose of the Study:
- To develop and validate a simple, sensitive, and eco-friendly method for determining six phthalate esters in physiological saline.
- To assess the presence of these phthalates in commercially available saline solutions.
Main Methods:
- Monolithic silica spin column extraction (MonoSpin-SPE) for sample pretreatment.
- Ultra-fast liquid chromatography-mass spectrometry (UFLC-MS) for separation and detection.
- Optimization of extraction and chromatographic conditions for linearity, sensitivity, and reproducibility.
Main Results:
- The method demonstrated good linearity (R² > 0.995) across specified ranges for six phthalate esters.
- Low limits of detection (0.02–0.9 μ/L) and quantification (0.08–2.7 μ/L) were achieved.
- High reproducibility (RSDs < 11.2%) and recovery rates (71.2%–107.3%) were obtained.
- Dimethyl-phthalate (DMP) and diethyl-phthalate (DEP) were detected in two brands of physiological saline.
Conclusions:
- MonoSpin-SPE combined with UFLC-MS is a suitable method for phthalate ester analysis in physiological saline.
- The presence of DMP and DEP in saline samples highlights the need for quality control in medical supplies.
Related Concept Videos
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Determining the pH of Salt Solutions
Extraction: Advanced Methods
NMR Spectroscopy: Spin–Spin Coupling
Spin–Spin Coupling: One-Bond Coupling
Spin–Spin Coupling Constant: Overview
Qualitatively, any spin plus-half nucleus polarizes the spins of its electrons to the minus-half state. Consequently, the paired electron in the hydrogen–carbon bond must...