Dual-energy CT: a phantom comparison of different platforms for abdominal imaging
Thorsten Sellerer1, Peter B Noël2,3, Manuel Patino4
1Chair for Biomedical Physics, Department of Physics and Munich School of BioEngineering, Technical University of Munich, James-Franck-Str. 1, 85748, Garching, Germany. thorsten.sellerer@tum.de.
Objectives:
Evaluation of imaging performance across dual-energy CT (DECT) platforms, including dual-layer CT (DLCT), rapid-kVp-switching CT (KVSCT) and dual-source CT (DSCT).
Methods:
A semi-anthropomorphic abdomen phantom was imaged on these DECT systems. Scans were repeated three times for CTDIvol levels of 10 mGy, 20 mGy, 30 mGy and different fat-simulating extension rings. Over the available range of virtual-monoenergetic images (VMI), noise as well as quantitative accuracy of hounsfield units (HU) and iodine concentrations were evaluated.
Results:
For all VMI levels, HU values could be determined with high accuracy compared to theoretical values. For KVSCT and DSCT, a noise increase was observed towards lower VMI levels. A patient-size dependent increase in the uncertainty of quantitative iodine concentrations is observed for all platforms. For a medium patient size the iodine concentration root-mean-square deviation at 20 mGy is 0.17 mg/ml (DLCT), 0.30 mg/ml (KVSCT) and 0.77mg/ml (DSCT).
Conclusion:
Noticeable performance differences are observed between investigated DECT systems. Iodine concentrations and VMI HUs are accurately determined across all DECT systems. KVSCT and DLCT deliver slightly more accurate iodine concentration values than DSCT for investigated scenarios. In DLCT, low-noise and high-image contrast at low VMI levels may help to increase diagnostic information in abdominal CT.
Key Points:
• Current dual-energy CT platforms provide accurate, reliable quantitative information. • Dual-energy CT cross-platform evaluation revealed noticeable performance differences between different systems. • Dual-layer CT offers constant noise levels over the complete energy range.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
What is Energy?
Free Energy
Cell Potential and Free Energy
Thermodynamics is the branch of physics dealing with the relationship between heat and other forms of energy. In an electrochemical cell, chemical energy is converted into electrical energy.
Thus, a link can be predicted between cell potential, free energy change, and the equilibrium constant for the reaction. Cell potential can also be measured as the oxidant or the reducing strength, and similar acid-base strength measures are reflected in equilibrium...