Comparison of Patient Skin Dose Evaluated Using Radiochromic Film and Dose Calculation Software
Florian Magnier1,2, Maxime Poulin3, Claire Van Ngoc Ty4
1Pôle d'Imagerie Diagnostique et de Radiologie Interventionnelle, CHU, 63003, Clermont-Ferrand, France. florian.magnier@clermont.unicancer.fr.
Purpose:
To compare, in an interventional radiology setting, peak skin doses (PSDs) delivered as calculated using a dedicated software tool and as measured using radiochromic film. To assess the utility of this dose calculation software tool in routine clinical practice.
Materials And Methods:
First, radiochromic films were positioned on the examination table in the back of an adult anthropomorphic phantom to measure PSD, and X-ray examinations were simulated. Then, films were again positioned in the patient's back for 59 thoracic or abdominopelvic endovascular interventions. The results obtained with the radiochromic films were taken as a reference and were statistically compared with those of the software.
Results:
With measured PSDs ranging from 100 to 7000 mGy, the median software-film difference was 8.5%. Lin's concordance coefficient was 0.98 [0.97; 0.99] (p < 0.001), meaning that concordance was excellent between the two methods. For the films where PSD exceeded 1000 mGy, the median difference in the measured value was 8.7% [- 1.3; 21.1], with a maximum discrepancy of 34%. Lin's concordance coefficient was 0.98 [0.96; 1] (p < 0.001), meaning that concordance was excellent between the two methods.
Conclusion:
Comparison between radiochromic films and the software tool showed that the software is a suitable tool for a simple and reliable estimation of PSD. The software seems to be a good alternative to films, whose use remains complex.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
Drug Dosing: Infants and Children