Pacemaker implantation rate after transcatheter aortic valve implantation with early and new-generation devices: a systematic review
Related Experiment Video
Updated: Jan 30, 2026

Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Philippe J van Rosendael1, Victoria Delgado1, Jeroen J Bax1
1Department of Cardiology, Leiden University Medical Center, Albinusdreef 2, 2333 ZA Leiden, The Netherlands.
Aims:
The incidence of new-onset conduction abnormalities requiring permanent pacemaker implantation (PPI) after transcatheter aortic valve implantation (TAVI) with new-generation prostheses remains debated. This systematic review analyses the incidence of PPI after TAVI with new-generation devices and evaluates the electrical, anatomical, and procedural factors associated with PPI. In addition, the incidence of PPI after TAVI with early generation prostheses was reviewed for comparison.
Methods And Results:
According to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses checklist, this systematic review screened original articles published between October 2010 and October 2017, reporting on the incidence of PPI after implantation of early and new-generation TAVI prostheses. Of the 1406 original articles identified in the first search for new-generation TAVI devices, 348 articles were examined for full text, and finally, 40 studies (n = 17 139) were included. The incidence of a PPI after the use of a new-generation TAVI prosthesis ranged between 2.3% and 36.1%. For balloon-expandable prostheses, the PPI rate remained low when using an early generation SAPIEN device (ranging between 2.3% and 28.2%), and with the new-generation SAPIEN 3 device, the PPI rate was between 4.0% and 24.0%. For self-expandable prostheses, the PPI rates were higher with the early generation CoreValve device (16.3-37.7%), and despite a reduction in PPI rates with the new Evolut R, the rates remained relatively higher (14.7-26.7%). When dividing the studies according to the highest (>26.0%) and the lowest (<12.1%) quintile of PPI rate, patients within the highest quintile were more frequently women when compared with the lowest quintile group (50.9% vs. 46.3%, P < 0.001). Pre-existent conduction abnormalities (electrical factor), calcification of the left ventricular outflow tract (anatomical factor), and balloon valvuloplasty and depth of implantation (procedural factors) were associated with increased risk of PPI.
Conclusion:
The rate of PPI after TAVI with new-generation devices is highly variable. Specific recommendations for implantation of each prosthesis, taking into consideration the presence of pre-existent conduction abnormalities and anatomical factors, may be needed to reduce the risk of PPI.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
Related Concept Videos
07:35Hyperglycemic Clamp and Hypoglycemic Clamp in Conscious Mice
11:05SECONDs Administration Guidelines: A Fast Tool to Assess Consciousness in Brain-injured Patients
07:37Assessment and Communication for People with Disorders of Consciousness
08:31Conscious and Non-conscious Representations of Emotional Faces in Asperger's Syndrome
05:41Metabolic Profile Analysis of Zebrafish Embryos
11:10Hyperinsulinemic-euglycemic Clamps in Conscious, Unrestrained Mice