Laboratory-determined Phosphorus Flux from Lake Sediments as a Measure of Internal Phosphorus Loading
Published on: March 6, 2014
A novel automatic flow method with direct-injection photometric detector for determination of dissolved reactive
Stanislawa Koronkiewicz1, Mihaela Trifescu2, Lech Smoczynski2
1Department of Chemistry, University of Warmia and Mazury, 10-957, Olsztyn, Poland. stankor@uwm.edu.pl.
Abstract:
The novel automatic flow system, direct-injection detector (DID) integrated with multi-pumping flow system (MPFS), dedicated for the photometric determination of orthophosphates in wastewater and freshwater samples is for the first time described. All reagents and the sample were injected simultaneously, in counter-current into the reaction-detection chamber by the system of specially selected for this purpose solenoid micro-pumps. The micro-pumps provided good precision and accuracy of the injected volumes. For the determination of orthophosphates, the molybdenum blue method was employed. The developed method can be used to detect orthophosphate in the range 0.1-12 mg L-1, with the repeatability (RSD) about 2.2% at 4 mg L-1 and a very high injection throughput of 120 injections h-1. It was possible to achieve a very small consumption of reagents (10 μL of ammonium molybdate and 10 μL of ascorbic acid) and sample (20 μL). The volume of generated waste was only 440 μL per analysis. The method has been successfully applied, giving a good accuracy, to determination of orthophosphates in complex matrix samples: treated wastewater, lake water and reference sample of groundwater. The developed system is compact, small in both size and weight, requires 12 V in supply voltage, which are desirable for truly portable equipment used in routine analysis. The simplicity of the system should result in its greater long-time reliability comparing to other flow methods previously described.
More Related Videos
10:53Concentration of Virus Particles from Environmental Water and Wastewater Samples Using Skimmed Milk Flocculation and Ultrafiltration
Published on: March 17, 2023
08:21Optimized Procedure for Determining the Adsorption of Phosphonates onto Granular Ferric Hydroxide using a Miniaturized Phosphorus Determination Method
Published on: May 18, 2018
Related Concept Videos
The Phosphorus Cycle
Automatic Processing and Automatic Social Behavior
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Experimental Determination of Chemical Formula
Convenience Sampling Method
Convenience sampling is a non-random method of sample selection; this method selects individuals that are easily accessible and may result in biased data. For example, a marketing...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...