Magnetic Resonance Imaging Quantification of Pulmonary Perfusion using Calibrated Arterial Spin Labeling
Published on: May 30, 2011
Three-dimensional arterial spin labeling imaging with a DANTE preparation pulse
Tsuyoshi Matsuda1, Hirohiko Kimura2, Hiroyuki Kabasawa3
1Division of Ultrahigh Field MRI, Institute for Biomedical Sciences, Iwate Medical University, 2-1-1, Nishitokuta, Yahaba, Shiwa-Gun, Iwate 028-3694, Japan.
Delays Alternating with Nutation for Tailored Excitation (DANTE) preparation pulses effectively suppress bright intravascular signals in arterial spin-labeled (ASL) imaging. This new method maintains signal uniformity, unlike existing techniques, offering improved ASL data quality.
Area of Science:
- Medical Imaging
- Magnetic Resonance Imaging
- Physiology
Background:
- Arterial spin-labeled (ASL) imaging can exhibit bright intravascular signal artifacts.
- Vascular suppression (VS) techniques aim to reduce these artifacts by dephasing labeled spins.
- Current VS methods like Motion-Sensitized Driven-Equilibrium (MSDE) can reduce signal uniformity and increase echo time.
Purpose of the Study:
- To compare the efficacy of Delays Alternating with Nutation for Tailored Excitation (DANTE) preparation pulses for VS in ASL imaging against the MSDE technique.
- To evaluate the impact of DANTE-prepared ASL on signal intensity distribution and artifact reduction.
Main Methods:
- ASL imaging was performed using both DANTE-prepared and MSDE VS sequences.
- Studies were conducted on a flow phantom with moving water and on human volunteers.
- Signal intensity, signal decay patterns, and inter-hemispheric differences were analyzed.
Main Results:
- Both DANTE and MSDE reduced bright intravascular signal artifacts in volunteer ASL images.
- Phantom studies showed similar signal decay patterns for moving water between the two methods.
- DANTE-prepared ASL demonstrated smaller right-left signal intensity differences compared to MSDE.
Conclusions:
- DANTE-prepared ASL effectively suppresses vascular signals while preserving signal intensity uniformity.
- DANTE offers a promising alternative for vessel suppression in ASL imaging, potentially improving image quality.
- This method addresses limitations of existing VS techniques in ASL MRI.
More Related Videos
Related Concept Videos
NMR Spectroscopy: Spin–Spin Coupling
Spin–Spin Coupling: One-Bond Coupling
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Spin–Spin Coupling Constant: Overview
Qualitatively, any spin plus-half nucleus polarizes the spins of its electrons to the minus-half state. Consequently, the paired electron in the hydrogen–carbon bond must...
Spin–Spin Coupling: Two-Bond Coupling (Geminal Coupling)
The central atom need not be NMR-active because its electrons are affected by the electron polarization of the spin-active atoms. However, spin information is transmitted less effectively than in one-bond coupling, and 2J values are usually weaker than 1J values. The energy of...