Corneal endothelial morphology and function after torsional and longitudinal ultrasound mode phacoemulsification
László Jr Módis1, Eszter Szalai1, Zsuzsa Flaskó1
1Department of Ophthalmology, Clinical Center, University of Debrecen, Debrecen, Hungary.
Purpose:
To study the endothelial cell morphology and corneal thickness changes after phacoemulsification by using the OZil torsional and longitudinal ultrasound techniques (Infiniti Vision System, Alcon Laboratories).
Setting:
Department of Ophthalmology, Clinical Center, University of Debrecen, Debrecen, Hungary.
Methods:
52 patients with cataract were randomly assigned to longitudinal ultrasound and torsional mode group. All surgeries were performed through a 2.2 mm clear corneal incision, the method employed being divide and conquer. The endothelial morphometry such as cell density (ECD), mean cell area, coefficient of variation of cell area, and central corneal thickness were examined with specular microscopy (EM-1000, Tomey) preoperatively and 4, 8 weeks postoperatively.
Results:
ECD values decreased significantly in both surgical groups (P < .001, repeated- mesures ANOVA), the postoperative endothelial cell loss was higher in the longitudinal ultrasound mode group (3.5% and 6.5%, at 4 and 8 weeks after surgery) than in the torsional group (3.3% and 5.5%, at 4 and 8 weeks after surgery), the difference not being significant between the two groups (P = .164 and P = .479, at 4 and 8 weeks after surgery, Mann-Whitney test). There was no statistically significant difference in any of the assessed parameters between the two surgical groups (P > .05). No significant correlation was found between the endothelial cell loss and the nucleus density.
Conclusions:
Both phacoemulsification techniques were safe and effective. The torsional handpiece performs oscillatory movements and delivers less energy into the eye than the longitudinal ultrasound technique, therefore providing more favorable energy and thermal safety profile.
Related Concept Videos
Torsion of Noncircular Members
Longitudinal Research
Torsional Pendulum
As long as the rigid body's angular displacement is small, its oscillation can be modeled as a linear angular oscillation. The amplitude of the oscillation is an angle. The role of mass is played...
Bending and Torsional Moments
The reaction developed in a structural element when subjected to an external force causes the element to bend. When a structural element bends upwards, it creates compressive normal forces on the top and tensile normal forces on the bottom, resulting in a couple that determines the bending...
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Longitudinal Studies