[Effectivity and Safety of a Modified Tip Design in Torsional Phacoemulsification]
Sabine Schmidt1, Sophie Hubich1, Jan Markus Vetter1
1Augenklinik Berlin-Marzahn GmbH.
Background:
Torsional mode phacoemulsification results in more effective fragmentation of the nucleus due to a different movement of the phacotip. In this clinical study, we investigated the influence of a modified tip design and active fluidics on the efficacy of phacoemulsification and safety for the corneal endothelium.
Material And Methods:
We conducted a prospective randomized 2 : 1 study in which 40 patients were operated on with the mini-flared Kelman Tip using the Infiniti® System (group 1), and 20 patients were operated on with the Intrepid® Balanced Tip and the Centurion® System. We analyzed the intraoperative cumulative dissipated energy and also the density of the corneal endothelium measured with an endothelial microscope (CEM 530, Nidek) pre- and postoperatively.
Results:
Both groups did not differ preoperatively in age, sex, axial length of the globe or corneal endothelium cell density nor cataract density (LOCS3). All surgeries were uneventful. The cumulative dissipated energy in group 1 (mini-flared Kelman tip, Infiniti System) was 38% higher than in group 2 (balanced tip, Centurion System; p < 0.05). The endothelial cell loss was 8% in group 1 and 10.3% in group 2 (p > 0.05). The cell size (polymegathism) increased in both groups significantly with + 37 µm in group 1 (p < 0.05) und + 54 µm in group 2 (p < 0.05). There was no statistically significant difference between both groups (p > 0.05). The number of hexagonal cells (pleomorphism) and corneal thickness did not differ in both groups either pre- nor postoperatively.
Conclusions:
Compared to torsional phacoemulsification with a mini-flared Kelman Tip and gravity fluidics, torsional phacoemulsification with a modified tip design and active fluidics is 38% more effective regarding the cumulative dissipated energy. Endothelial cell loss occurs to a similar extend using both systems. The postoperative changes in cell size (polymegathism), number of hexagonal cells (pleomorphism) and corneal thickness (pachymetry) were similar among both systems. We conclude, that the intraoperative stress on the endothelium is equivalent with both systems used.
Related Concept Videos
Torsion of Noncircular Members
Torsional Pendulum
As long as the rigid body's angular displacement is small, its oscillation can be modeled as a linear angular oscillation. The amplitude of the oscillation is an angle. The role of mass is played...
Bending and Torsional Moments
The reaction developed in a structural element when subjected to an external force causes the element to bend. When a structural element bends upwards, it creates compressive normal forces on the top and tensile normal forces on the bottom, resulting in a couple that determines the bending...
Survey Safety
Group Design
Factorial Design