Intrapatient study comparing 3D printed bolus versus standard vinyl gel sheet bolus for postmastectomy chest wall
James L Robar1, Kathryn Moran2, James Allan2
1Department of Radiation Oncology, Dalhousie University, Halifax, Canada; Department of Physics and Atmospheric Science, Dalhousie University, Halifax, Canada; Nova Scotia Health Authority, Halifax, Canada.
Purpose:
This patient study evaluated the use of 3-dimensional (3D) printed bolus for chest wall radiation therapy compared with standard sheet bolus with regard to accuracy of fit, surface dose measured in vivo, and efficiency of patient setup. By alternating bolus type over the course of therapy, each patient served as her own control.
Methods And Materials:
For 16 patients undergoing chest wall radiation therapy, a custom 5.0 mm thick bolus was designed based on the treatment planning computed tomography scan and 3D printed using polylactic acid. Cone beam computed tomography scanning was used to image and quantify the accuracy of fit of the 2 bolus types with regard to air gaps between the bolus and skin. As a quality assurance measure for the 3D printed bolus, optically stimulated luminescent dosimetry provided in vivo comparison of surface dose at 7 points on the chest wall. Durations of patient setup and image guidance were recorded and compared.
Results:
In 13 of 16 patients, the bolus was printed without user intervention, and the median print time was 12.6 hours. The accuracy of fit of the bolus to the chest wall was improved significantly relative to standard sheet bolus, with the frequency of air gaps 5 mm or greater reduced from 30% to 13% (P < .001) and maximum air gap dimension diminished from 0.5 ± 0.3 to 0.3 ± 0.3 mm on average. Surface dose was within 3% for both standard sheet and 3D printed bolus. On average, the use of 3D printed bolus reduced the setup time from 104 to 76 seconds.
Conclusions:
This study demonstrates 3D printed bolus in postmastectomy radiation therapy improves fit of the bolus and reduces patient setup time marginally compared with standard vinyl gel sheet bolus. The time savings on patient setup must be weighed against the considerable time needed for the 3D printing process.
Related Concept Videos
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Nonlinear Pharmacokinetics: Drug Elimination for IV Bolus Injection
Following the administration of a single intravenous (IV) bolus injection, we can determine the concentration of the drug in the plasma at any given time. This calculation is achieved using a specific equation that integrates the values of Vmax and KM.
We can also...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution
Biological Effects of Radiation