Background:
Estimated continuous cardiac output (esCCOTM) based on pulse wave transit time is one of alternative non-invasive CO measurement techniques.
Methods:
Randomized study included 23 scheduled patients operated upon due to cardiovascular diseases. Cardiac index (CI) was measured Comparative analyses of esCCO and others CO measurement methods used intraoperative was carried out. In the first group (n = 9) esCCO was compared with transpulmonary thermodilution (PiCCO-plus); in the second group (n = 8) - with pulmonary artery thermodilution; in the third group (n = 6) - with transoesophageal echocardiography (velocity-time integral).
Results:
In the 1st group direct correlation was found (r = 0,773, p <0.0001), but overestimation was found in 39% of cases; underestimation in 4%. The 2nd group showed direct correlation (r = 0.586, p <0.0001). The 3d group showed direct relationship (r = 0.68, p = 0.0018), but 66.7% of the measurements were out of reference interval (more than ? 15%). Blend- Altman method showed the dispersion of results in all groups.
Conclusions:
1. Estimated continuous cardiac output measurement technique based on PWTT has a direct correla- tion with prepulmonary thermodilution and transoesophageal echocardiography, medium and high power respectively. 2. esCCO has significant differences with the referential techniques during general anesthesia in cardiac surgery pa- tients. 3. Calibration based on invasive blood pressure and outside cardiac output measurement does not increase the accuracy of measurements. 4. esCCO has a negative diagnostic value and cannot be recommendedfor the cardiac out- put evaluation during cardiac surgery. 5. This method can be useful for analyze general effectiveness of perioperative hemodynamics.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Special considerations while measuring pulse