In Vitro Evaluation of ECG-Synchronized Pulsatile Flow Using the i-cor Diagonal Pump in Neonatal and Pediatric ECLS
Morgan Moroi1, Madison Force1, Shigang Wang1
1Department of Pediatrics, Penn State Health Pediatric Cardiovascular Research Center, Penn State College of Medicine, Penn State Health Children's Hospital, Hershey, PA, USA.
Abstract:
The objective was to assess the i-cor electrocardiogram-synchronized diagonal pump in terms of hemodynamic energy properties for off-label use in neonatal and pediatric extracorporeal life support (ECLS) circuits. The neonatal circuit consisted of an i-cor pump and console, a Medos Hilite 800 LT oxygenator, an 8Fr arterial cannula, a 10Fr venous cannula, 91 cm of 0.6-cm ID arterial tubing, and 91 cm of 0.6-cm ID venous tubing. The pediatric circuit was identical except it included a 12Fr arterial cannula, a 14Fr venous cannula, and a Medos Hilite 2400 LT oxygenator. Neonatal trials were conducted at 36°C with hematocrit 40% using varying flow rates (200-600 mL/min, 200 mL increments) and postarterial cannula pressures (40-100 mm Hg, 20 mm Hg increments) under nonpulsatile mode and pulsatile mode with various pulsatile amplitudes (1000-4000 rpm, 1000 rpm increments). Pediatric trials were conducted at different flow rates (800-1600 mL/min, 400 mL/min increments). Mean pressure and energy equivalent pressure increased with increasing postarterial cannula pressure, flow rate, and pulsatile amplitude. Physiologic-like pulsatility was achieved between pulsatile amplitudes of 2000-3000 rpm. Pressure drops were greatest across the arterial cannula. Pulsatile flow generated significantly higher total hemodynamic energy (THE) levels than nonpulsatile flow. THE levels at postarterial cannula site increased with increasing postarterial cannula pressure, pulsatile amplitude, and flow rate. No surplus hemodynamic energy (SHE) was generated under nonpulsatile mode. Under pulsatile mode, preoxygenator SHE increased with increasing postarterial cannula pressure and pulsatile amplitude, but decreased with increasing flow rate. The i-cor system can provide nonpulsatile and pulsatile flow for neonatal and pediatric ECLS. Pulsatile amplitudes of 2000-3000 rpm are recommended for use in neonatal and pediatric patients.
More Related Videos
Related Concept Videos
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
ATP Driven Pumps II: P-type Pumps
A typical P-type pump has three cytosolic domains: nucleotide-binding (N), phosphorylation (P), and activator (A) domains. These domains are connected to the membrane-spanning helices by short amino acid segments. ATP hydrolysis and covalent phosphoenzyme intermediate formation are crucial parts of the catalytic cycle. At the highly...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Pumped Concrete
For direct-acting pumps, the concrete enters the pump via the inlet valve under the action of gravity and suction created by the movement of the piston. This concrete is then forced into the pipeline and out through the outlet valve by the forward movement...
Refrigerators and Heat Pumps
A household refrigerator removes heat from...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...