Subcutaneous ICD screening with the Boston Scientific ZOOM programmer versus a 12-lead ECG machine
Shu C Chang1, Kristen K Patton2, Melissa R Robinson2
1Division of Cardiology, University of Colorado, Aurora, Colorado, USA.
Background:
The subcutaneous implantable cardioverter-defibrillator (S-ICD) requires preimplant screening to ensure appropriate sensing and reduce risk of inappropriate shocks. Screening can be performed using either an ICD programmer or a 12-lead electrocardiogram (ECG) machine. It is unclear whether differences in signal filtering and digital sampling change the screening success rate.
Methods:
Subjects were recruited if they had a transvenous single-lead ICD without pacing requirements or were candidates for a new ICD. Screening was performed using both a Boston Scientific ZOOM programmer (Marlborough, MA, USA) and General Electric MAC 5000 ECG machine (Fairfield, CT, USA). A pass was defined as having at least one lead that fit within the screening template in both supine and sitting positions.
Results:
A total of 69 subjects were included and 27 sets of ECG leads had differing screening results between the two machines (7%). Of these sets, 22 (81%) passed using the ECG machine but failed using the programmer and five (19%) passed using the ECG machine but failed using the programmer (P < 0.001). Four subjects (6%) passed screening using the ECG machine but failed using the programmer. No subject passed screening with the programmer but failed with the ECG machine.
Conclusion:
There can be occasional disagreement in S-ICD patient screening between an ICD programmer and ECG machine, all of whom passed with the ECG machine but failed using the programmer. On a per lead basis, the ECG machine passes more subjects. It is unknown what the inappropriate shock rate would be if an S-ICD was implanted. Clinical judgment should be used in borderline cases.
Related Concept Videos
The Scientific Method
Generally, predictions are tested using carefully-designed experiments. Based on the outcome of these...
The Scientific Method
The Scientific Method
Scientific Laws and Theories
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Machines
A free-body diagram of the...