A flow-leak correction algorithm for pneumotachographic work-of-breathing measurement during high-flow nasal cannula
Francesco Montecchia1, Fabio Midulla2, Paola Papoff3
1Medical Engineering Laboratory, Department of Civil Engineering and Computer Science Engineering, University of Rome "Tor Vergata", Rome, Italy.
Abstract:
Measuring work of breathing (WOB) is an intricate task during high-flow nasal cannula (HFNC) therapy because the continuous unidirectional flow toward the patient makes pneumotachography technically difficult to use. We implemented a new method for measuring WOB based on a differential pneumotachography (DP) system, equipped with one pneumotachograph inserted in the HFNC circuit and another connected to a monitoring facemask, combined with a leak correction algorithm (LCA) that corrects flow measurement errors arising from leakage around the monitoring facemask. To test this system, we used a mechanical lung model that provided data to compare LCA-corrected respiratory flow, volume and time values with effective values obtained with a third pneumotachograph used instead of the LCA to measure mask flow leaks directly. Effective and corrected volume and time data showed high agreement (Bland-Altman plots) even at the highest leak. Studies on two healthy adult volunteers confirmed that corrected respiratory flow combined with esophageal pressure measurements can accurately determine WOB (relative error < 1%). We conclude that during HFNC therapy, a DP system combined with a facemask and an algorithm that corrects errors due to flow leakages allows pneumotachography to measure reliably the respiratory flow and volume data needed for calculating WOB.
More Related Videos
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Gene Flow
Breathing
Flow Cytometry
In...
Flow Sheet
Here's a closer look at the examples of flowsheets commonly used by nurses:
Graphic Sheet Documentation: