Plugger temperature of cordless heat carriers according to the time elapsed
Hoon-Sang Chang1, Se-Hee Park2, Kyung-Mo Cho2
1Department of Conservative Dentistry, Chonnam National University School of Dentistry, Gwangju, Korea.
Objective:
The purpose of this study was to measure the temperature of the plugger tip of 3 cordless heat carriers set at 200°C.
Materials And Methods:
Pluggers of the same taper (0.06, 0.08, 0.10) and similar tip sizes (sizes of 50 and 55) from 3 cordless heat carriers, namely SuperEndo-α2 (B & L Biotech), Friendo (DXM), and Dia-Pen (Diadent), were used and an electric heat carrier, System B (SybronEndo), was used as the control. The plugger tips were covered with customized copper sleeves, heated for 10 seconds, and the temperature was recorded with a computerized measurement system attached to a K-type thermometer at room temperature (n = 10). The data were analyzed with 2-way analysis of variance at a 5% level of significance.
Results:
The peak temperature of the plugger tips was significantly affected by the plugger taper and by the heat carrier brand (p < 0.05). The peak temperature of the plugger tips was between 177°C and 325°C. The temperature peaked at 207°C-231°C for the 0.06 taper pluggers, 195°C-313°C for the 0.08 taper pluggers, and 177°C-325°C for the 0.10 taper pluggers. Only 5 of the 12 plugger tips showed a temperature of 200°C ± 10°C. The time required to reach the highest temperature or 200°C ± 10°C was at least 4 seconds.
Conclusion:
When using cordless heat carriers, clinicians should pay attention to the temperature setting and to the activation time needed to reach the intended temperature of the pluggers.
More Related Videos
06:13Single-Cell Proteomics Preparation for Mass Spectrometry Analysis Using Freeze-Heat Lysis and an Isobaric Carrier
Published on: December 9, 2022
08:52Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere
Published on: April 30, 2018
Related Concept Videos
Electron Carriers
Over the many stages of cellular respiration, glucose breaks down into carbon dioxide and water. Electron carriers pick up electrons lost by glucose in these reactions, temporarily storing and releasing them into the electron...
Quantifying Heat
Specific Heat
For example, increasing the temperature of one gram of water by 1°C requires one calorie of heat energy and can be written as 1 cal/g-°C, or...
Carrier Transport
Drift Current:
The drift of charge carriers is started by an external electric field (E). Charged particles, such as electrons and holes, experience an acceleration between collisions with lattice atoms. For electrons, this results in a drift velocity (vd) given by:
Heating and Cooling Curves
For instance, the addition of heat raises the temperature of a solid; the amount of heat absorbed depends on the heat capacity of the solid (q = mcsolidΔT). According to thermochemistry, the relation between the amount of heat absorbed or released by a substance, q, and its...
Heat Flow and Specific Heat