Contrast media injection protocol optimization for dual-energy coronary CT angiography: results from a circulation
Domenico De Santis1,2, Damiano Caruso2, U Joseph Schoepf3
1Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Drive, Charleston, SC, USA.
Objectives:
To investigate the minimum iodine delivery rate (IDR) required to achieve diagnostic coronary attenuation (300 HU) with dual-energy coronary CTA.
Methods:
Acquisitions were performed on a circulation phantom with a third- generation dual-source CT scanner. Contrast media was injected for a fixed time whilst IDRs varied from 1.0 to 0.3 gI/s in 0.1-gI/s intervals. Noise-optimized virtual monoenergetic imaging (VMI+) reconstructions from 40 to 90 keV in 5 keV increments were generated. Contrast-to-noise ratio (CNR) and coronary HU were measured for each injection.
Results:
VMI+ from 40-70 keV reached diagnostic attenuation with at least one IDR. The minimum IDR achieving a diagnostic attenuation ranged from 0.4 gI/s at 40 keV (312.8 HU) to 1.0 gI/s at 70 keV (334.1 HU). Attenuation values reached with IDR of 1.0 gI/s were significantly higher at each keV level (p<0.001). CNR showed a near perfect correlation with the IDR (ρ≥0.962; p<0.001), the IDR of 1.0 gI/s provided the highest CNR at each keV level, achieving the highest overall value at 40 keV (54.0±3.1).
Conclusions:
IDRs from 0.4-1.0 gI/s associated with VMI+ from 40-70 keV provide diagnostic coronary attenuation with dual-energy coronary CTA.
Key Points:
• Iodine delivery rate (IDR) is a major determinant of contrast enhancement. • Low-keV noise-optimized monoenergetic images (VMI+) maximize iodine attenuation. • Low-keV VMI+ allows for lower IDRs while maintaining adequate coronary attenuation. • Lowest IDR to reach 300 HU was 0.4 gI/s, 40 keV VMI+.
More Related Videos
Related Concept Videos
Coronary Circulation
Coronary circulation begins at the base of the aorta, where two main arteries arise—the left and right coronary arteries. These arteries encircle the heart in the coronary sulcus and supply the...
Optimal Foraging
What is Energy?
Free Energy
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...