Instrumentation and signal processing for the detection of heavy water using off axis-integrated cavity output
A Gupta1, P J Singh1, D Y Gaikwad2
1Atomic and Molecular Physics Division, Bhabha Atomic Research Centre, Mumbai 400085, India.
Abstract:
An experimental setup is developed for the trace level detection of heavy water (HDO) using the off axis-integrated cavity output spectroscopy technique. The absorption spectrum of water samples is recorded in the spectral range of 7190.7 cm-1-7191.5 cm-1 with the diode laser as the light source. From the recorded water vapor absorption spectrum, the heavy water concentration is determined from the HDO and water line. The effect of cavity gain nonlinearity with per pass absorption is studied. The signal processing and data fitting procedure is devised to obtain linear calibration curves by including nonlinear cavity gain effects into the calculation. Initial calibration of mirror reflectivity is performed by measurements on the natural water sample. The signal processing and data fitting method has been validated by the measurement of the HDO concentration in water samples over a wide range from 20 ppm to 2280 ppm showing a linear calibration curve. The average measurement time is about 30 s. The experimental technique presented in this paper could be applied for the development of a portable instrument for the fast measurement of water isotopic composition in heavy water plants and for the detection of heavy water leak in pressurized heavy water reactors.
Related Concept Videos
Regulation of Water Output
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Emission Spectroscopy: Instrumentation
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
¹H NMR Signal Integration: Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....