Related Experiment Video
Updated: Feb 13, 2026

In Vivo Modeling of the Morbid Human Genome using Danio rerio
Published on: August 24, 2013
Comparison of different types of phacoemulsification tips. III. Morphological changes induced after multiple uses in
Konstantinos T Tsaousis1, David F Chang1, Liliana Werner1
1From the John A. Moran Eye Center (Tsaousis, Werner, Guan, Reiter, Li, Mamalis) and the Utah Nanofab (Perez), College of Engineering, University of Utah, Salt Lake City, Utah, and the Altos Eye Physicians (Chang), Los Altos, California, USA.
Purpose:
To assess the ultrastructural morphologic changes in the surface's micro-roughness for single-use and multiple-use phaco tips after multiple phacoemulsification cycles in an ex vivo animal model.
Setting:
John A. Moran Eye Center and Utah Nanofab, College of Engineering, University of Utah, Salt Lake City, Utah, USA.
Design:
Experimental study.
Methods:
Eight types of phaco tips were studied using an identical experimental protocol. Two tips of each type were used to emulsify porcine cataractous lenses of moderate hardness for 2 minutes. This cycle was then repeated 5 consecutive times for each phaco tip. Scanning electron microscopy and white-light interferometry were used to analyze each tip for potential ultrastructural damage.
Results:
No significant structural damage was noted on either single- or multiple-use tips after 5 cycles of experimental use. There was a trend toward increased surface micro-roughness at the tip opening after multiple procedures. Most tips had salt crystals on their surface and deposits of organic material composed of carbon and oxygen, which were consistent with porcine tissue that had not been cleaned from the tip surface.
Conclusions:
No significant ultrastructural damage was detected after repeated experimental use of phaco tips in a porcine cataract model. There was no significant difference when phaco tips labeled for single use or multiple uses were compared.
Related Concept Videos
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Sources of Self-Esteem III: Social Comparison
Types of Functions III
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
Types of Reports III: Telephone and Verbal Reports
Here's an overview of each type:
Telephone Orders
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.