Accuracy of left atrial bipolar voltages obtained by ConfiDENSE multielectrode mapping in patients with persistent
Baptiste Maille1, Moloy Das2, Ahmed Hussein1
1Institute of Cardiovascular Medicine and Science, Department of Cardiology, Liverpool Heart and Chest Hospital, Thomas Drive, Liverpool, L14 3PE, UK.
Introduction:
The ConfiDENSE™ module (Carto3 v4) allows rapid annotation of endocardial electrograms acquired by multielectrode (ME) mapping. However, its accuracy in assessing atrial voltages is unknown.
Methods And Results:
Two ConfiDENSE™ left atrial voltage maps were created during continuous pacing in 20 patients undergoing catheter ablation for persistent AF using a ME lasso catheter and a contact force (CF) sensing ablation catheter. The automated tissue proximity indicator (TPI) filter was then applied to the ME map to yield a TPI map. Reference maps (RM) were created offline by a blinded observer by manually assessing all points against fidelity criteria. Bipolar voltages and proportion of low voltage points (< 0.5 mV) derived from the ME, CF, and TPI maps were compared with those derived from the RM. Note that 853 ± 365 points, 252 ± 184 points, and 144 ± 73 were collected for ME, TPI, and CF maps, respectively, and 429 ± 153 points were included in the RM. Voltages with CF and TPI maps were similar to those with RM (1.57 ± 0.47 mV vs. 1.63 ± 0.31 mV, P = 0.57 and 1.50 ± 0.38 mV vs. 1.63 ± 0.31 mV, P = 0.07, respectively), whereas ME maps showed a significantly lower mean voltage (1.00 ± 0.22 mV, P < 0.001). As compared to RM maps (17 ± 8%), low voltage points were significantly overestimated by the ME maps (50 ± 9% (P < 0.001) and TPI maps (28 ± 13% (P < 0.001), but not by the CF maps (22 ± 14%, P = 0.17).
Conclusion:
Application of the TPI filter to ConfiDENSE maps significantly increases the quality of the voltage data, conserving a reasonable point density, but still overestimates low voltage points as compared to CF-sensing maps or maps reviewed manually.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Amyloid Fibrils
Amyloid deposits were observed as early as 1639 in the liver and the spleen. In 1854, Rudolph Virchow performed iodine staining,...
Amyloid Fibrils
Uncertainty in Measurement: Accuracy and Precision
Improving Translational Accuracy
Bipolar Disorder