Lessons Learned from Implantation of Morcher 50D and 96S Artificial Iris Diaphragms
1Stein Eye Institute, David Geffen School of Medicine, University of California, Los Angeles, California, USA.
Purpose:
To discuss problems associated with the implantation of two Morcher iris diaphragm models.
Methods:
We describe the history, intraoperative complications, and postoperative complications of 5 patients with specific Morcher iris implants.
Results:
We implanted Morcher 50D devices in 1 patient and Morcher 96S devices in 4 patients. Complications included postoperative rotation, device mis-sizing, difficult intraoperative rotation, zonular dehiscence, and intraoperative hemorrhage.
Conclusion:
Artificial iris implantation has a steep learning curve. With widespread availability on the horizon in the United States, the sharing of surgical experiences is key to achieving the best outcomes for patients.
Related Concept Videos
Avoidance Learning and Learned Helplessness
Avoidance learning occurs when an organism learns that a specific behavior can prevent an unpleasant outcome. For example, a student who receives a bad grade may start studying harder to avoid future poor grades. This behavior persists even when the negative outcome is no longer present. Avoidance learning is powerful because it maintains behavior in the absence of the...