Assessing abrasion of orthodontic surface sealants using a modified ophthalmic optical coherence tomography device
Sinan Şen1, Ralf Erber2, Kevin Kunzmann3
1Department of Orthodontics and Dentofacial Orthopaedics, University of Heidelberg, Heidelberg, Germany. sinan.sen@med.uni-heidelberg.de.
Objective:
Optical coherence tomography (OCT) is a clinical standard in ophthalmology. Currently, its application in dentistry is gaining increasing interest. In this study, we tested the possibility to use a modified commercially available spectral domain OCT (SD-OCT) to assess the layer thickness of orthodontic surface sealants.
Materials And Methods:
Reference samples of surface sealants for calibration and repeatability testing were measured using a micrometer screw. SD-OCT measurements were compared with micro-CT and light microscopic analyses. After validating the calibration of the SD-OCT, surface sealant layer thickness after aging (thermo cycling) and simulation of professional tooth cleaning (PTC) was assessed using the SD-OCT on 45 extracted teeth assigned to three test groups (n = 15 each): Light Bond™ Sealant, Pro Seal®, and Opal® Seal.
Results:
SD-OCT showed excellent repeatability and accuracy for measurements of surface sealant layer thickness. Compared with micro-CT, SD-OCT showed better accordance with the reference measurements. The analysis of surface sealants after thermo cycling and PTC revealed poor resistance of Light Bond after only aging and demonstrated substantial wear of all sealants after aging and PTC.
Conclusion:
Imaging using commercially available ophthalmic SD-OCT might represent a suitable non-invasive methodology for longitudinal assessments of surface sealant layer thickness in vitro and in vivo.
Clinical Relevance:
SD-OCT might be a suitable non-invasive method for longitudinal assessments of surface sealant durability in clinical trials.
Related Concept Videos
Abrasion Resistance of Concrete
One such test is the revolving disc test, where three plates...
Ophthalmic Drug Delivery Systems
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Surface Tension and Surface Energy
Consider a beaker filled with liquid. The bulk molecules in the liquid experience equal attractive forces on all sides with the surrounding molecules. However, the surface molecules experience a net attractive force downward due to the bulk molecules. The surface of the liquid behaves like a stretched membrane,...
Modified Boxplots
However, the box plot does not tell the reader about outliers - values that lie far from the center of the data. We can modify the standard box and whisker plot to identify the outliers and visualize the actual spread of the data in a sample.
Initially, we calculate the adjusted...
Mnemonic Devices
Acronyms
Acronyms are created by using the initial letters of a series of words to form a new word or phrase. This approach condenses complex information into a single, memorable entity. For example,...