Background:
Early enteral nutrition has been shown to decrease complications and improve patient outcomes. Post pyloric feeding is recommended for patients with gastric intolerance or at high risk for aspiration. Feeding tube placement can be challenging and pose risk of pulmonary complications. Reliance on radiographic confirmation for feeding tube placement exposes the patient to radiation. Electromagnetic placement device (EMPD) may offer a method to minimize pulmonary complications, increase successful placement, and decrease radiation exposure to the patient.
Objective:
The purpose of this study was to evaluate the safety and efficacy of using EMPD verification, instead of routine abdominal radiographic confirmation, for small-bore feeding tube placement.
Results:
Variables evaluated were adverse events, utilization of radiographs for confirmation, and success rate of feeding tube placement in the ordered location. Two time frames were reviewed. In a 1-year period, 3754 small-bore feeding tubes were placed using EMPD, with zero adverse events noted. Radiographic confirmation was utilized in 0%-29.2% of the EMPD placed tubes. Successful placement of feeding tubes using EMPD ranged from 94%-99.6%. During a 5-year period, 7081 EMPD feeding tubes were evaluated. One adverse event, pneumothorax, occurred during the placement of these 7081 tubes, for a rate of 0.014%.
Conclusion:
Feeding tube placement confirmation is safe and efficacious via EMPD providing an effective method of feeding tube placement with a success rate >94% into the desired location. EMPD is an accurate verification method of distal tip location, eliminating the need for routine abdominal radiographic confirmation.
More Related Videos
05:11High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition
Published on: June 27, 2025
06:37Using Reference Reagents to Confirm Robustness of Cytokine Release Assays for the Prediction of Monoclonal Antibody Safety
Published on: September 15, 2023
Related Concept Videos
Confirmation Biases
The Electromagnetic Spectrum
The Electromagnetic Spectrum
Survey Safety
Self-Efficacy
Electromagnetic Waves