Minimizing a Wireless Passive LC-Tank Sensor to Monitor Bladder Pressure: A Simulation Study
Jacob Melgaard1, Johannes J Struijk1, Nico J M Rijkhoff1
1Department of Health Science and Technology, The Faculty of Medicine, Aalborg University, Fredrik Bajers Vej 7, 9220 Aalborg Ø, Denmark.
Abstract:
In this simulation study, a wireless passive LC-tank sensor system was characterized. Given the application of continuous bladder monitoring, a specific system was proposed in terms of coil geometries and electronic circuitry. Coupling coefficients were spatially mapped by simulation, as a function of both coil distance, and longitudinal and transverse translation of the sensor relative to the antenna. Further, two interrogation schemes were outlined. One was an auto-balancing bridge for computing the sensor-system impedance. In this case, the theoretical noise limit of the analogue part of the system was found by simulations. As the full system is not necessary for obtaining a pressure reading from the sensor, a simplified circuit more suited for an implantable system was deduced. For this system, both the analogue and digital parts were simulated. First, the required ADC resolution for operating the system at a given coupling was found by simulations in the noise-free case. Then, for one selected typical operational point, noise was added gradually, and through Monte-Carlo type simulations, the system performance was obtained. Combining these results, it was found that it at least is possible to operate the proposed system for distances up to 12 mm, or equivalently for coupling coefficients above 0.005. In this case a 14 bit ADC is required, and a carrier SNR of 27 dB can be tolerated.
Related Concept Videos
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Errors occurring during blood pressure monitoring
Several factors...
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Active versus Passive Immunity
Active Immunity
Active immunity refers to the resistance one develops...
Urinary Bladder
In males, the bladder is situated in front of the rectum, while in females, it is positioned anterior to the vagina and uterus. The bladder floor contains an inverted triangular area called the trigone, defined by the two ureteric...
Passive Diffusion: Overview and Kinetics
When administered orally, drugs establish a substantial concentration gradient between the gastrointestinal (GI) lumen and the bloodstream, expediting...