Electronic collimation of radiographic images: does it comprise an overexposure risk?
11 Medical Physics Department, General Anticancer Oncology Hospital of Athens 'Agios Savvas' , Athens , Greece.
Objective:
To investigate whether electronic collimation software, which is available in all digital X-ray systems, may comprise an overexposure risk.
Methods:
In the context of surveys on Diagnostic Reference Levels carried out in two radiographic facilities, along with data on exposure factors, the radiographic field sizes were also recorded. In one facility (Unit A), a wireless flat panel detector is used with a conventional X-ray unit, while in the other, a fully digital system is installed (Unit B). The electronically collimated image sizes were compared with the original radiation field sizes. The differences between these two systems concerning the field sizes and the mode of electronic collimation utilization were investigated.
Results:
In Unit A, manual electronic collimation was extensively used and cases where the radiation field size was up to three times larger than that electronically collimated, were identified. On the contrary, in Unit B radiation fields were smaller and electronic collimation was automatic.
Conclusion:
When electronic collimation is used in manual mode instead of proper pre-exposure collimation, then it does comprise an overexposure risk. The risk is larger in radiographic units where the field size is not automatically selected according to the examination protocol and no interlocks against oversized collimation settings exist. Advances in knowledge: When radiologists review masked images to make the diagnosis, possible suboptimal X-ray field collimation practices may go unnoticed for long. Therefore, radiologists and medical physicists should periodically survey the original images to determine the actual radiation field sizes used for each radiographic examination type.
Related Concept Videos
Relative Risk
Electron Carriers
Over the many stages of cellular respiration, glucose breaks down into carbon dioxide and water. Electron carriers pick up electrons lost by glucose in these reactions, temporarily storing and releasing them into the electron...
Electron Affinity
Factors Affecting the Risk of Infection
The integrity and count of the white blood cells help the body resist pathogens and fight infection. When impaired, it reduces the body's resistance to pathogens. The acidic pH levels of the gastrointestinal, genitourinary tracts, and skin...
Electron Behavior
Electrons are negatively charged subatomic particles that are attracted to an orbit around the positively-charged nucleus of an atom. They reside in locations that are associated with energy levels called shells and are further organized into sub-shells and orbitals within each shell.
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the...
Electron Transport Chains
The ETC is comprised of...