Related Experiment Video
Updated: Jan 15, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Exposure index in digital radiography and its dependence on acquisition parameters, anatomy, and manufacturer
Ioannis A Tsalafoutas1, Shady AlKhazzam1, Mohammed Hassan Kharita1
1Medical Physics Section, OHS Department, Hamad Medical Corporation, Doha, Qatar.
Purpose:
The exposure index (EI), the target exposure index (EIT), and the deviation index (DI) have been defined in the IEC Standard 62494-1 Ed.1 2008-08. This study investigates the impact of certain acquisition parameters, the imaged anatomy, and the manufacturer's specificities on the EI of radiological images and how these may affect EIT setting procedure.
Methods:
Images were acquired using two digital radiography (DR) systems of two different manufacturers, using aluminum attenuators and an anthropomorphic phantom. Acquisition parameters like the tube potential (kVp), the tube loading (mAs), the exposure time, the automatic exposure control (AEC) system settings (sensor and dose level selection), the grid (with or without), the additional filtration, the field size, and the imaged anatomy were varied and their effect on the EI was quantified separately for each system.
Results:
EI is linearly related to the incident air kerma (IAK) on the detector as expected (by definition). For constant IAK, EI increases with increasing kVp. While EI in general is reduced in the presence of scatter, this may not always be the case. Under AEC operation, even the exposure time can make a difference. EI is strongly affected by the imaged anatomy in combination with the AEC sensor and field size selections, the examination protocol, and the manufacturer.
Conclusions:
Many parameters affect the EI calculation apart from IAK. Among them, the most important are the imaged anatomy and the manufacturer. Since the EI calculation is a complex procedure, setting of the EIT values should be done with caution on a per-examination and manufacturer basis, since the values that apply for one digital system are not always applicable to another. Furthermore, when EI is used as an image quality tool, a DI variation of at least ±2 should be allowed before a possibly meaningful red flag is activated.
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Biological Effects of Radiation
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...

