Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
A new sampling algorithm demonstrates that ultrasound equipment cleanliness can be improved.
Greg S Whiteley1, Trevor O Glasbey2, Sue C Westerway3
1Whiteley Corporation, North Sydney, Australia; Western Sydney University, School of Medicine, Sydney, NSW, Australia.
A study found 26% of patient-ready ultrasound equipment in 5 facilities lacked cleanliness. A cleaning intervention significantly improved standards, showing adenosine triphosphate (ATP) testing is effective for monitoring probe cleanliness.
Area of Science:
- Healthcare-associated infections
- Medical device reprocessing
- Diagnostic imaging technology
Background:
- Australian guidelines exist for cleaning reusable ultrasound probes and equipment.
- This preliminary study assessed the cleanliness of patient-ready ultrasound equipment across five healthcare facilities in a major city.
Purpose of the Study:
- To investigate the cleanliness standards of ultrasound equipment in clinical use.
- To evaluate the effectiveness of adenosine triphosphate (ATP) testing as a quality assurance tool for equipment cleanliness.
Main Methods:
- Adenosine triphosphate (ATP) testing with a sampling algorithm was used to assess surface cleanliness, measuring relative light units (RLUs).
- Surfaces were tested for compliance with literature-recommended cleanliness levels (<100 RLUs).
- A cleaning intervention step (CIS) using disposable wipes was performed, followed by retesting for ATP levels.
Main Results:
- Out of 253 tested surfaces, 26% (66/253) showed inadequate cleanliness.
- Following the cleaning intervention, over 91% (135/148) of surfaces demonstrated significant improvement in cleaning standards (P>.001).
- For 6% (9/148) of surfaces, the cleaning intervention needed repeating to achieve target cleanliness (<25 RLUs).
Conclusions:
- Adenosine triphosphate (ATP) testing serves as an effective, real-time quality assurance method for monitoring ultrasound probe and equipment cleanliness.
- The study highlights the need for consistent adherence to cleaning protocols to ensure patient safety.
More Related Videos
05:18Author Spotlight: Ultrasound-Guided Needle Release Combined with Corticosteroid Injection for the Treatment of Supinator Syndrome
Published on: May 26, 2023
10:19Substernal Thyroid Biopsy Using Endobronchial Ultrasound-guided Transbronchial Needle Aspiration
Published on: November 10, 2014
Related Concept Videos
Leveling Equipment
Trial and Error and Algorithm
Personal Protective Equipment
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Handwashing I: Introduction and Types of Equipment
Hand wash basins in clinical areas should have faucets that can be turned on and off without using the hands; that is, they should be non-touch or lever-operated....