The Influence of Corneal Biomechanical Properties on Intraocular Pressure Measurements Using a Rebound Self-tonometer
Lyndsay Brown1, William Foulsham2, Savva Pronin1
1Princess Alexandra Eye Pavilion, Department of Ophthalmology, University of Edinburgh, Edinburgh, UK.
Purpose:
The purpose of this study was to examine the effect of corneal biomechanical properties on intraocular pressure (IOP) measurements obtained using a rebound self-tonometer (Icare HOME) compared with Goldmann applanation tonometry (GAT).
Methods:
An observational study of 100 patients with glaucoma or ocular hypertension. All had a comprehensive ophthalmic examination and standard automated perimetry. IOP was assessed by GAT, Icare HOME and Ocular Response Analyzer, which was also used to assess corneal hysteresis (CH) and corneal resistance factor (CRF). Central corneal thickness (CCT) was recorded.
Results:
Mean (±SD) IOP measurements were 14.3±3.9 and 11.7±4.7 mm Hg using GAT and Icare HOME, respectively. Average CCT, CRF, and CH were 534.5±37.3 μm, 9.0±1.7 mm Hg, and 9.4±1.5 mm Hg, respectively. The mean difference between Icare HOME and GAT was -2.66±3.13 mm Hg, with 95% limits of agreement of -8.80 to 3.48 mm Hg, however, there was evidence of proportional bias. There was negative correlation between IOP and CH [5.17 mm Hg higher Icare HOME IOP (P=0.041, R=0.029) and 7.23 mm Hg higher GAT IOP (P=0.008, R=0.080) for each 10 mm Hg lower CH], whereas thinner CCT was significantly associated with lower IOP (P<0.001, R=0.14 for Icare HOME and P<0.001, R=0.08 for GAT). In multivariable analysis, although CRF and CH remained associated with IOP measured using either GAT or Icare HOME, CCT was no longer significant.
Conclusion:
IOP measurements obtained using a self-tonometer, similar to GAT, were more influenced by overall corneal biomechanics than CCT.
Related Concept Videos
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Vapor Pressure Lowering
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.