Related Experiment Video
Updated: Feb 13, 2026

Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
AzBio Speech Understanding Performance in Quiet and Noise in High Performing Cochlear Implant Users
Jason A Brant1, Steven J Eliades1, Hannah Kaufman1
1Department of Otorhinolaryngology-Head and Neck Surgery, Hospital of the University of Pennsylvania.
Objective:
To evaluate high-performing cochlear implant patients' performance on AzBio sentence testing.
Methods:
Retrospective review of prospectively collected database at a tertiary care hospital. Unilateral cochlear implant patients with AzBio testing were included. The primary outcome of interest was AzBio performance scores in quiet and at +10 and +5 decibels signal to noise (dB S/N).
Results:
One hundred eighty five subjects met inclusion criteria with scores for AzBio in quiet, 114 at +10 dB S/N, and 66 at +5 dB S/N. Linear mixed effects models showed performance significantly correlated with time since activation in all conditions (8.4%/yr; p < 0.0001). Strong correlations between mean performance in quiet and at +10 dB S/N (r = 0.77, p < 0.0001), and between +10 and +5 dB S/N (r = 0.73, p < 0.0001) were found. The correlation between quiet and +5 dB S/N (r = 0.45, p = 0.01) was less robust. Shapiro-Wilks test of normality found only +10 dB S/N to correspond to a normal distribution. Skew analysis demonstrated values of -0.64, -0.11, and 0.8 for quiet, +10 dB S/N, and +5 dB S/N, respectively.
Conclusions:
AzBio scores at +10 dB S/N show a strong correlation with, but avoid the ceiling effects that limit the usefulness of testing in quiet for high performing cochlear implant users, making it the preferred test in this population. Significant complexities exist in the study of outcomes in cochlear implant recipients, and there is no single test that is ideal for the entirety of this population.
Related Concept Videos
Pathophysiology of Cardiac Performance
Exercise and Muscle Performance
Endurance exercises
Endurance exercises involve running, swimming, or cycling, which require repetitive movements with low force output. When a person engages in endurance exercise, a few noticeable changes occur in their skeletal muscles. For instance, the number of capillaries...
Sources of Self-Esteem II: Performance Feedback
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Elution Process