Integrated Photoacoustic Ophthalmoscopy and Spectral-domain Optical Coherence Tomography
Published on: January 15, 2013
SENSITIVITY AND SPECIFICITY OF DETECTING POLYPOIDAL CHOROIDAL VASCULOPATHY WITH EN FACE OPTICAL COHERENCE TOMOGRAPHY
Talisa E de Carlo1,2,3,4,5, Gregg T Kokame2,3,4,5,6, Kyle N Kaneko3,4,5
1Department of Medicine, University of Hawaii Transitional Residency Program, University of Hawaii School of Medicine, Honolulu, Hawaii.
Purpose:
Determine sensitivity and specificity of polypoidal choroidal vasculopathy (PCV) diagnosis with structural en face optical coherence tomography (OCT) and OCT angiography (OCTA).
Methods:
Retrospective review of the medical records of eyes diagnosed with PCV by indocyanine green angiography with review of diagnostic testing with structural en face OCT and OCTA by a trained reader. Structural en face OCT, cross-sectional OCT angiograms alone, and OCTA in its entirety were reviewed blinded to the findings of indocyanine green angiography and each other to determine if they could demonstrate the PCV complex. Sensitivity and specificity of PCV diagnosis was determined for each imaging technique using indocyanine green angiography as the ground truth.
Results:
Sensitivity and specificity of structural en face OCT were 30.0% and 85.7%, of OCT angiograms alone were 26.8% and 96.8%, and of the entire OCTA were 43.9% and 87.1%, respectively. Sensitivity and specificity were improved for OCT angiograms and OCTA when looking at images taken within 1 month of PCV diagnosis.
Conclusion:
Sensitivity of detecting PCV was low using structural en face OCT and OCTA but specificity was high. Indocyanine green angiography remains the gold standard for PCV detection.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...
Properties of Enantiomers and Optical Activity
Imaging Studies III: Computed Tomography
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...