Four-Minute Bone SPECT Using Large-Field Cadmium-Zinc-Telluride Camera
Purpose:
The aim of this study was to determine the minimum acquisition time without decreasing lesion detectability of bone SPECT using a whole-body cadmium-zinc-telluride camera.
Methods:
Patients referred for bone SPECT were retrospectively included. SPECT of 30 patients were reframed from native data (16 s/projection) to produce 10-, 5-, and 3-s/projection data sets. A "critical" acquisition time/projection was defined as that below which the SPECT quality becomes insufficient for interpretation, as determined by 3 reviewers using a 4-point scale (0 = quality insufficient for interpretation, 1 = average, 2 = good, 3 = excellent). Three reviewers (blinded to the acquisition time) evaluated SPECT data sets (n = 79), native and reframed with "critical" acquisition times, in a randomized order. A lesion was defined as any uptake considered pathological by a reviewer. Lesion detectability equivalence between native SPECT and reframed SPECT was assessed by calculating a coefficient (κ) for each reviewer.
Results:
Image quality of the first sample (n = 30) was significantly and progressively less well graded for the reframed data sets by all reviewers. Only 1 patient was graded 0 by each reviewer for the 5-s/projection data set. For the 3-s/projection data set, 3 patients were graded 0. No patients were graded 0 for 10-s/projection data set. The minimal acquisition time, for each projection, was defined as 5 s/projection. The coefficient κ, between native and reframed, with critical acquisition time/projection SPECT was greater than 0.9 for each reviewer.
Conclusions:
The more contrasted images of the cadmium-zinc-telluride camera allow performance of 5-s/projection SPECT without loss of lesion detectability. This suggests the possibility of performing whole-body SPECT in a reasonable time or reducing injected doses, especially in pediatric patients.
Related Concept Videos
Field Effect Transistor
Electric Field
In the new picture, imagine that the first charge sets up an electric field independent of all other charges in the universe. When another charge comes in its vicinity, the second charge experiences an electric force depending on the electric field at that point. The source charge does not...
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Electromagnetic Fields
However, the observation of...
Bone Structure
Bone Remodeling