An automated Monte Carlo QC system for volumetric modulated arc therapy: Possibilities and challenges
R Chakarova1, R Cronholm2, M Krantz3
1Department of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Gothenburg, Sweden; Department of Radiation Physics, Sahlgrenska Academy at the University of Gothenburg, Sweden.
Purpose:
To develop and implement an automated Monte Carlo (MC) system for patient specific VMAT quality control in a patient geometry that generates treatment planning system (TPS) compliant DICOM objects and includes a module for 3D analysis of dose deviations. Also, the aims were to recommend diagnose specific tolerance criteria and an evaluation procedure.
Methods:
The EGSnrc code package formed the basis for development of the MC system. The workflow consists of a number of modules connected to a TPS by means of manual DICOM exports and imports which were executed sequentially without user interaction. DVH comparison was performed in the TPS. In addition, MC- and TPS dose distributions were analysed by applying the normalized dose difference (NDD) formalism. NDD failure maps and a pass rate for a certain threshold were obtained. 170 clinical plans (prostate, thorax, head-and-neck and gynecological) were selected for analysis.
Results:
Agreement within 1.5% was found between clinical- and MC data for the mean dose to the target volumes and within 3% for parameters more sensitive to the shape of the DVH e.g. D98% PTV. Regarding the NDD analysis, tolerance criteria 2%/3 mm were established for prostate plans and 3%/3 mm for the rest of the cases.
Conclusions:
An automated MC system was developed and implemented. Evaluation procedure is recommended with NDD-analysis as a first step. For pass rate < 95%, the evaluation continues with comparison of DVH parameters. For deviations larger than 2%, a visual inspection of the clinical- and MC dose distributions is performed.
More Related Videos
Related Concept Videos
Arc Length of a Curve
Arc Length Function
Arc Length of a Curve: Problem Solving
Gene Therapy
Group Therapy
Behavior Therapy
Exposure therapy is a cornerstone of behavioral treatment for anxiety disorders. It involves systematic exposure to feared stimuli, either in real...
![Automated Preparation of [68Ga]Ga-3BP-3940 on a Synthesis Module for PET Imaging of the Tumor Microenvironment](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F68356.jpg&w=3840&q=50)