Tissue oximetry by diffusive reflective visible light spectroscopy: Comparison of algorithms and their robustness
Nassim Nasseri1,2, Stefan Kleiser1, Ursula Wolf2
1Department of Neonatology, University Hospital Zurich, Zurich, Switzerland.
Abstract:
It is essential to measure tissue oxygen saturation (StO2 ) locally and in thin layers of tissue, for example, the bronchial mucosa, skin flaps and small bones. Visible light spectroscopy (VLS) with a shallow penetration depth is suitable method. Although several VLS algorithms have been developed and described, they have not yet been compared to each other. This hinders attempts to compare the clinical results obtained by different algorithms. To address this issue, we compared the algorithms of Harrison, Knoefel, Pittman-Duling, Sato and our OxyVLS oximeter, which applies the algorithm from Wodick and Lübbers, in a liquid phantom with optical properties of human tissue. We generally observed considerable differences between the algorithms, which were StO2 dependent. Exceptions were OxyVLS and Sato, showing a high level of agreement with negligible StO2 dependency. In spite of the considerable deviation between the other algorithms, the difference of StO2 between them in clinically normal StO2 was <10%. We did not observe any dependency of the algorithms on hemoglobin content of the phantom or temperature.
More Related Videos
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Diffusion
Diffusion