Maladjustment of programmable ventricular shunt valves by inadvertent exposure to a common hospital device
R Fujimura1, R Lober1, K Kamian1
1Department of Neurosurgery, Dayton Children's Hospital, Wright State Boonshoft School of Medicine, Dayton, Ohio, United States.
Background:
Programmable ventricular shunt valves are commonly used to treat hydrocephalus. They can be adjusted to allow for varying amounts of cerebrospinal fluid (CSF) flow using an external magnetic programming device, and are susceptible to maladjustment from inadvertent exposure to magnetic fields.
Case Description:
We describe the case of a 3-month-old girl treated for hydrocephalus with a programmable StrataTM II valve found at the incorrect setting on multiple occasions during her hospitalization despite frequent reprogramming and surveillance. We found that the Vocera badge, a common hands-free wireless communication system worn by our nursing staff, had a strong enough magnetic field to unintentionally change the shunt setting. The device is worn on the chest bringing it into close proximity to the shunt valve when care providers hold the baby, resulting in the maladjustment.
Conclusion:
Some commonly used medical devices have a magnetic field strong enough to alter programmable shunt valve settings. Here, we report that the magnetic field from the Vocera hands-free wireless communication system, combined with the worn position, results in shunt maladjustment for the StrataTM II valve. Healthcare facilities using the Vocera badges need to put protocols in place and properly educate staff members to ensure the safety of patients with StrataTM II valves.
Related Concept Videos
Hospitals-II
Nurses that work in...
Hospitals-I
Shunt Admittances
To model this effect, the method of images is employed. This method involves replacing the Earth with an image conductor that mirrors the original...
Common Ion Effect
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Common Respiratory Disorders
Upper respiratory disorders impact the airways above the vocal cords, encompassing areas like the nose, sinuses, and throat. Various conditions fall under this category, including the common cold and allergic rhinitis. These disorders can stem from several causes,...