Impact of intraoperative mechanical ventilation on left ventricular lead function in cardiac resynchronization
Christian von Bary1, Viktoria Laski1, Hannes Fricke1
1Medizinische Klinik I, Rotkreuzklinikum München, Akademisches Lehrkrankenhaus der Technischen Universität München, München, Germany.
Background:
Intraoperative controlled mechanical ventilation (CMV) changes the intrathoracic geometry and may impact postoperative left ventricular (LV) lead function after CRT implantation. This multicenter study investigates the effect of intraoperative ventilation setting (spontaneous breathing [SB] vs CMV) on postoperative LV lead function taking into account patients' body mass index (BMI).
Methods:
CRT implantation was performed at two centers during SB in 92 and during CMV in 73 patients. Follow-up was carried out after 3 ± 5 and 36 ± 53 days. Functional lead parameters (FLP; pacing threshold and impedance), postoperative adverse events (A; phrenic nerve stimulation [PNS] and lead malfunction), and patients' BMI were assessed. Delta values of FLP between baseline and follow-up visits were analyzed applying an analysis of covariance model to detect subclinical alterations in LV lead function.
Results:
AE occurred in a total of 36 (21%) patients. PNS was observed in 26 (15%) patients and LV lead repositioning due to malfunction was necessary in 10 (6%) patients. Both AE and FLP delta values between baseline and follow-up were not associated with intraoperative ventilation settings nor the patients' BMI.
Conclusions:
This study demonstrates that there is no impact of the intraoperative ventilation setting (SB vs CMV) on postoperative FLP or the occurrence of AE. This is also the case taking into account the BMI. With respect to these findings both approaches-sedation only or general anaesthesia including CMV-can be safely implemented during CRT implantation.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Mechanical Protein Functions
Mechanism of Cardiac Arrhythmias
Cardiac Catheterization III: Left Heart Catheterization