Related Experiment Videos
Sensitometric control of roentgen film processors
1Department of Medical Engineering, Karolinska Institutet, Stockholm, Sweden.
Acta Radiologica (Stockholm, Sweden : 1987)
|May 1, 1987
Abstract:
Monitoring of film processors performance is essential since image quality, patient dose and costs are influenced by the performance. A system for sensitometric constancy control of film processors and their associated components is described. Experience with the system for 3 years is given when implemented on 17 film processors. Modern high quality film processors have a stability that makes a test frequency of once a week sufficient to maintain adequate image quality. The test system is so sensitive that corrective actions almost invariably have been taken before any technical problem degraded the image quality to a visible degree.