Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield
Abbas Shokri1, Leila Ramezani1, Mohsen Bidgoli2
1Dental Implant Research Center, Department of Oral and Maxillofacial Radiology, Faculty of Dentistry, Hamadan University of Medical Sciences, Hamadan, Iran.
Purpose:
This study aimed to evaluate the effect of field-of-view (FOV) size on the gray values derived from conebeam computed tomography (CBCT) compared with the Hounsfield unit values from multidetector computed tomography (MDCT) scans as the gold standard.
Materials And Methods:
A radiographic phantom was designed with 4 acrylic cylinders. One cylinder was filled with distilled water, and the other 3 were filled with 3 types of bone substitute: namely, Nanobone, Cenobone, and Cerabone. The phantom was scanned with 2 CBCT systems using 2 different FOV sizes, and 1 MDCT system was used as the gold standard. The mean gray values (MGVs) of each cylinder were calculated in each imaging protocol.
Results:
In both CBCT systems, significant differences were noted in the MGVs of all materials between the 2 FOV sizes (P<.05) except for Cerabone in the Cranex3D system. Significant differences were found in the MGVs of each material compared with the others in both FOV sizes for each CBCT system. No significant difference was seen between the Cranex3D CBCT system and the MDCT system in the MGVs of bone substitutes on images obtained with a small FOV.
Conclusion:
The size of the FOV significantly changed the MGVs of all bone substitutes, except for Cerabone in the Cranex3D system. Both CBCT systems had the ability to distinguish the 3 types of bone substitutes based on a comparison of their MGVs. The Cranex3D CBCT system used with a small FOV had a significant correlation with MDCT results.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Biodiversity and Human Values
Professional Values
The values that are the foundation of the nursing profession are altruism, autonomy, human dignity, and social justice.
First, altruism refers to the concern for the welfare and well-being of others without personal...
Critical Values
z Scores and Unusual Values
This score indicates how far a value is from the mean in terms of standard deviation. For example, if a data value has a z score of +1, the researcher can infer that the particular data value is one standard deviation above the mean. If another data...