Related Experiment Video
Updated: Feb 12, 2026

A Quantitative Fitness Analysis Workflow
Published on: August 13, 2012
Chromatographic characterisation of 11 phytocannabinoids: Quantitative and fit-to-purpose performance as a function
Matthew Noestheden1,2, Gareth Friedlander3, Jason Anspach3
1Chemistry Department, University of British Columbia Okanagan, Kelowna, BC, Canada.
Introduction:
Cannabis sativa L. (cannabis) is utilised as a therapeutic and recreational drug. With the legalisation of cannabis in many countries and the anticipated regulation of potency that will accompany legalisation, analytical testing facilities will require a broadly applicable, quantitative, high throughput method to meet increased demand. Current analytical methods for the biologically active components of cannabis (phytocannabinoids) suffer from low throughput and/or an incomplete complement of relevant phytocannabinoids.
Objective:
To develop a rapid, quantitative and broadly applicable liquid chromatography-tandem mass spectrometry analytical method for 11 phytocannabinoids in cannabis with acidic and neutral character.
Methodology:
Bulk diffusion coefficients were calculated using the Taylor-Aris open tubular method, with four reference compounds used to validate the experimental set-up. Three columns were quantitatively evaluated using van Deemter plots and fit-to-purpose performance metrics. Low (1.2 μL2 ) and standard (3.6 μL2 ) extra-column variance ultra-high pressure liquid chromatography (UPLC) configurations were contrasted. Method performance was demonstrated with methanolic cannabis flower extracts.
Results:
Bulk diffusion coefficients and van Deemter plots for 11 phytocannabinoids are reported. The developed chromatographic method includes the challenging Δ8 /Δ9 -tetrahydrocannabinol isobars and, at 6.5 min, is faster than existing methods targeting similar panels of biologically active phytocannabinoids.
Conclusions:
The bulk diffusion coefficients and van Deemter curves informed the development of a rapid quantitative method and will facilitate potential expansion to include additional compounds, including synthetic cannabinoids. The developed method can be implemented with low or standard extra-column variance UPLC configurations.
Related Concept Videos
Variance
The standard deviation measures the spread in the same units as the data....
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Induced-fit Model
Enzymes exhibit substrate specificity, meaning that they can only bind to certain substrates. This is mainly determined by the shape and chemical...
Purposive Learning
Inclusive Fitness
Chromatographic Methods: Terminology