Impact of a New Adaptive Statistical Iterative Reconstruction (ASIR)-V Algorithm on Image Quality in Coronary
Gianluca Pontone1, Giuseppe Muscogiuri1, Daniele Andreini2
1Centro Cardiologico Monzino, IRCCS, Via C. Parea 4, 20138, Milan, Italy.
Rationale And Objectives:
A new postprocessing algorithm named adaptive statistical iterative reconstruction (ASIR)-V has been recently introduced. The aim of this article was to analyze the impact of ASIR-V algorithm on signal, noise, and image quality of coronary computed tomography angiography.
Materials And Methods:
Fifty consecutive patients underwent clinically indicated coronary computed tomography angiography (Revolution CT; GE Healthcare, Milwaukee, WI). Images were reconstructed using filtered back projection and ASIR-V 0%, and a combination of filtered back projection and ASIR-V 20%-80% and ASIR-V 100%. Image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were calculated for left main coronary artery (LM), left anterior descending artery (LAD), left circumflex artery (LCX), and right coronary artery (RCA) and were compared between the different postprocessing algorithms used. Similarly a four-point Likert image quality score of coronary segments was graded for each dataset and compared. A cutoff value of P < .05 was considered statistically significant.
Results:
Compared to ASIR-V 0%, ASIR-V 100% demonstrated a significant reduction of image noise in all coronaries (P < .01). Compared to ASIR-V 0%, SNR was significantly higher with ASIR-V 60% in LM (P < .01), LAD (P < .05), LCX (P < .05), and RCA (P < .01). Compared to ASIR-V 0%, CNR for ASIR-V ≥60% was significantly improved in LM (P < .01), LAD (P < .05), and RCA (P < .01), whereas LCX demonstrated a significant improvement with ASIR-V ≥80%. ASIR-V 60% had significantly better Likert image quality scores compared to ASIR-V 0% in segment-, vessel-, and patient-based analyses (P < .01).
Conclusions:
Reconstruction with ASIR-V 60% provides the optimal balance between image noise, SNR, CNR, and image quality.
More Related Videos
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
10:16Retrograde Perfusion and Filling of Mouse Coronary Vasculature as Preparation for Micro Computed Tomography Imaging
Published on: February 10, 2012
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Statistical Significance
Probability in Statistics
An example of a simple event is a coin toss. The result of a coin toss is either a head or a tail. Here, head and tail are two simple events. These two simple events make up the sample space. Further, the probability of an event occurring falls within the range of 0 to 1. The probability of an...
Introduction to Statistics
In statistics, the collection of individuals or objects under study is called population. The idea of sampling is to select a portion of the larger population...