Characterization of OSL dosimeters for use in dose assessment in Computed Tomography procedures
Louise Giansante1, Josilene C Santos1, Nancy K Umisedo1
1Nuclear Physics Department, Institute of Physics, University of São Paulo, São Paulo, Brazil.
Abstract:
This study describes the characterization of an Al2O3:C OSLD (Landauer's Luxel™ tape) for dose evaluation in Computed Tomography. The irradiations were conducted using both a constant potential X-ray equipment and a 64-slice clinical CT scanner, and the readouts were performed using a Risø TL/OSL reader. The following aspects were studied: batch homogeneity, energy response, linearity of dose response, reproducibility, reusability, and effect of uncertainties with the normalization of OSL signals per their response to beta radiation. A group of 330 dosimeters from the 452 irradiated with the same dose presented OSL signals within the interval of 4.7% from the average. The dosimeters presented energy-dependent response in good agreement with results found in the literature. The air kerma response of the OSL signal showed a linear trend for both the constant potential X-ray device and the clinical CT scanner, with differences in their slopes of approximately 10%. Reproducibility, reusability, and effect of beta normalization were analyzed by separating 72 dosimeters in 3 groups. The results obtained in this study together with those of previous works indicate that this type of dosimeter is adequate for dose evaluation in CT clinical applications.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
Pre-Procedural Guidelines for Assessing Blood Pressure