How Well Does Dual-Energy Computed Tomography With Metal Artifact Reduction Software Improve Image Quality and
Shingo Ohira, Naoyuki Kanayama, Kentaro Wada
1Medical Physics and Engineering, and.
Objective:
The objective of this study was to assess the accuracy of the quantitative measurements obtained using dual-energy computed tomography with metal artifact reduction software (MARS).
Methods:
Dual-energy computed tomography scans (fast kV-switching) are performed on a phantom, by varying the number of metal rods (Ti and Pb) and reference iodine materials. Objective and subjective image analyses are performed on retroreconstructed virtual monochromatic images (VMIs) (VMI at 70 keV).
Results:
The maximum artifact indices for VMI-Ti and VMI-Pb (5 metal rods) with MARS (without MARS) were 17.4 (166.7) and 34.6 (810.6), respectively; MARS significantly improved the mean subjective 5-point score (P < 0.05). The maximum differences between the measured Hounsfield unit and theoretical values for 5 mg/mL iodine and 2-mm core rods were -42.2% and -68.5%, for VMI-Ti and VMI-Pb (5 metal rods), respectively, and the corresponding differences in the iodine concentration were -64.7% and -73.0%, respectively.
Conclusions:
Metal artifact reduction software improved the objective and subjective image quality; however, the quantitative values were underestimated.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Design Example: Traverse Angle Computations
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Area Computation by the Alternative Coordinate Method
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...