Recording characteristics of electrical impedance-electromyography needle electrodes
H Kwon1, J F Di Cristina2, S B Rutkove1
1Department of Neurology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA 02215-5491, United States of America.
Objective:
Needle EMG remains the standard clinical test for neuromuscular disease (NMD) assessment, but it only characterizes myofiber membrane depolarization. On the other hand, electrical impedance provides non-electrically active structural and compositional data of tissues. Here, we designed a prototype of needle electrode integrating electrical impedance and EMG measurement capabilities, the so-called I-EMG needle electrode.
Approach:
We use finite element method models to study the impedance recording characteristics of I-EMG needle electrodes. The simulated electrical and mechanical design specifications are then manufactured to create a prototype of an I-EMG needle electrode. We pilot these new needle electrodes by conducting in vivo impedance measurements with muscle at rest on healthy wild-type (wt, n = 5) and muscular dystrophy (mdx, n = 5) mice. Comparisons between wt and mdx mice are performed using Mann-Whitney test, two-tailed, p < 0.05. The electrical characterization of the EMG electrode in the developed I-EMG needles was performed in vitro on saline solution and through EMG detection in wt animal at rest and during voluntary contractions.
Results:
Muscle impedance demonstrate good repeatability (p < 0.05 and p < 0.005 for resistance and reactance at 50 kHz, respectively) and agreement between different I-EMG needles. Impedance data allows us to discriminate between mdx and wt muscle (p < 0.05 and p < 0.005 for resistance and reactance at 10 kHz, respectively). EMG broadband noise power and peak amplitude using the I-EMG needle were similar to that of a commercial monopolar EMG needle. EMG recordings using the I-EMG needle measured electrical activity similar to a standard monopolar needle with muscle at rest and during voluntary contraction.
Significance:
Needle I-EMG technology may offer the opportunity to enhance the diagnostic capability and quantification of NMD beyond that possible with either impedance or EMG techniques separately. Ultimately, needle I-EMG could serve as a new bedside tool to assess NMD without increasing the complexity or duration of the EMG test.
More Related Videos
09:36In Situ Lithiated Reference Electrode: Four Electrode Design for In-operando Impedance Spectroscopy
Published on: September 12, 2018
12:30Electric Cell-substrate Impedance Sensing for the Quantification of Endothelial Proliferation, Barrier Function, and Motility
Published on: March 28, 2014
Related Concept Videos
Characteristics of Life
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Standard Electrode Potentials
Characteristics of Fluids
Characteristics of BJT
For input characteristics, the base-emitter voltage is varied, maintaining a constant collector-emitter voltage. This setup reveals a Shockley-type dependence of the collector current on...