Scanning electron microscopic characteristics of manually loaded and preloaded foldable acrylic intraocular lenses
Mahmoud Fathy Rateb1, Zeiad Hasan Eldaly1, Walid Saad Eldein Ibrahim1
11 Department of Ophthalmology, Assiut University Hospitals, Assiut, Egypt.
Purpose::
To compare surface alterations between preloaded and manually loaded intraocular lens.
Methods::
Scanning electron microscope was utilized to evaluate surface alteration and deposits in four different types of intraocular lenses: preloaded hydrophobic acrylic, preloaded hydrophilic acrylic, manually loaded hydrophobic acrylic, and manually loaded hydrophilic acrylic. Six lenses with different powers (+6 D, +22 D, and +29 D) were used from each category, to represent different thickness categories of the intraocular lenses.
Results::
In total, 30 intraocular lenses have been evaluated in this study: 4 from the control group (2 hydrophobic and 2 hydrophilic lenses) and 12 from the preloaded intraocular lens and manually loaded groups (6 hydrophilic and 6 hydrophobic lenses with different powers). Surface deposits were found in eight hydrophobic intraocular lenses compared to a single intraocular lens with scattered deposits on the optical surface of a hydrophilic intraocular lens. In manually loaded intraocular lens group, five hydrophobic and one hydrophilic intraocular lenses showed identifiable marks on the optical surface. In the preloaded intraocular lens group, three hydrophobic intraocular lenses showed identifiable marks on the optical surface and three hydrophobic intraocular lenses showed surface wrinkling. All hydrophilic intraocular lenses revealed no identifiable marks.
Conclusion::
Surface alterations and deposits are a common finding in both preloaded and manually loaded intraocular lenses. Water content of acrylic intraocular lenses is an important factor predisposing to these changes, and hydrophobic intraocular lenses are more vulnerable than hydrophilic lenses. The impact on the final visual outcome needs further studies.
More Related Videos
11:14Comprehensive Characterization of Extended Defects in Semiconductor Materials by a Scanning Electron Microscope
Published on: May 28, 2016
03:07Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Related Concept Videos
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Characteristics of Life
Leaky Scanning
Electron Carriers
Over the many stages of cellular respiration, glucose breaks down into carbon dioxide and water. Electron carriers pick up electrons lost by glucose in these reactions, temporarily storing and releasing them into the electron...
Characteristics of Fluids