Wide scan imaging with swept-source optical coherent tomography for glaucoma diagnosis
Eun Hee Hong1, Yong Un Shin1, Min Ho Kang1
1Department of Ophthalmology, Hanyang University College of Medicine, Seoul, Korea.
Wide scans using swept-source optical coherence tomography (SS-OCT) effectively detect glaucoma. A single wide scan provides comparable diagnostic ability to standard macula and disc scans, simplifying glaucoma evaluation.
Area of Science:
- Ophthalmology
- Medical Imaging
- Glaucoma Diagnosis
Background:
- Glaucoma diagnosis relies on detecting structural changes in the retina.
- Optical coherence tomography (OCT) is crucial for measuring retinal nerve fiber layer (RNFL) and macular ganglion cell complex (GCC) thickness.
- Differentiating between glaucoma and healthy eyes requires accurate and repeatable measurements.
Purpose of the Study:
- To compare the glaucoma-discriminating capabilities of wide-field (12x9mm) SS-OCT macular and circumpapillary RNFL (cpRNFL) thickness measurements against standard (6x6mm) scans.
- To assess if a single wide scan can replace multiple standard scans for glaucoma evaluation.
Main Methods:
- Retrospective chart review of 122 subjects (60 glaucoma, 62 healthy).
- Comparison of macular ganglion cell layer plus inner plexiform layer (mGCIPL), mGCIPL plus nerve fiber layer (mGCC), and total retinal layer (TRL) thickness from wide and standard macula scans.
- Assessment of cpRNFL thickness from wide and standard disc scans using SS-OCT.
- Evaluation of measurement repeatability and agreement using intraclass correlation coefficients (ICCs).
- Analysis of glaucoma-discriminating ability using areas under the receiver operating characteristic curves (AUCs).
Main Results:
- All parameters demonstrated high repeatability and agreement (ICCs ≥ 0.800).
- AUCs for mGCIPL thickness ranged from 0.701-0.847, and for cpRNFL thickness ranged from 0.726-0.902 across scan types.
- No significant differences in AUCs were observed between wide and standard scans for either macular or cpRNFL measurements.
Conclusions:
- Wide SS-OCT scans show excellent agreement with standard scans for mGCIPL, mGCC, and cpRNFL measurements.
- The glaucoma-discriminating ability of wide scans is comparable to that of standard macula and disc scans.
- A single wide scan can effectively replace separate standard macula and disc scans in glaucoma assessment, streamlining the diagnostic process.
Related Concept Videos
Glaucoma: Overview
Leaky Scanning
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Imaging Studies III: Computed Tomography
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET