Safety Studies for a 44-Channel Suprachoroidal Retinal Prosthesis: A Chronic Passive Study
Carla J Abbott1,2, David A X Nayagam3,4, Chi D Luu1,2
1Centre for Eye Research Australia, Royal Victorian Eye and Ear Hospital, East Melbourne, Australia.
Purpose:
Following successful clinical outcomes of the prototype suprachoroidal retinal prosthesis, Bionic Vision Australia has developed an upgraded 44-channel suprachoroidal retinal prosthesis to provide a wider field of view and more phosphenes. The aim was to evaluate the preclinical passive safety characteristics of the upgraded electrode array.
Methods:
Ten normal-sighted felines were unilaterally implanted with an array containing platinum electrodes (44 stimulating and 2 returns) on a silicone carrier near the area centralis. Clinical assessments (color fundus photos, optical coherence tomography, full-field electroretinography, intraocular pressure) were performed under anesthesia prior to surgery, and longitudinally for up to 20 weeks. Histopathology grading of fibrosis and inflammation was performed in two animals at 13 to 15 weeks.
Results:
Eight animals showed safe electrode array insertion (good retinal health) and good conformability of the array to the retinal curvature. Eight animals demonstrated good mechanical stability of the array with only minor (<2 disc diameters) lateral movement. Four cases of surgical or stability complications occurred due to (1) bulged choroid during surgery, (2) hemorrhage from a systemic bleeding disorder, (3) infection, and (4) partial erosion of thin posterior sclera. There was no change in retinal structure or function (other than that seen at surgery) at endpoint. Histopathology showed a mild foreign body response. Electrodes were intact on electrode array removal.
Conclusions:
The 44-channel suprachoroidal electrode array has an acceptable passive safety profile to proceed to clinical trial. The safety profile is expected to improve in human studies, as the complications seen are specific to limitations (anatomic differences) with the feline model.
More Related Videos
10:50Computational Modeling of Retinal Neurons for Visual Prosthesis Research - Fundamental Approaches
Published on: June 21, 2022
10:00Techniques for Processing Eyes Implanted with a Retinal Prosthesis for Localized Histopathological Analysis: Part 2 Epiretinal Implants with Retinal Tacks
Published on: February 14, 2015
Related Concept Videos
Case Studies
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Survey Safety
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Ion Channels
Ion channels are specialized integral membrane proteins on the plasma membrane that allow...
Active versus Passive Immunity
Active Immunity
Active immunity refers to the resistance one develops...