A Comparative Pulse Accuracy Study of Two Commercially Available Patch Insulin Infusion Pumps
Jenna L Bowen1, Chris J Allender1
1School of Pharmacy and Pharmaceutical Sciences, Cardiff University, Cardiff, UK.
This study compared the accuracy of two insulin patch pumps. Results show significant differences in pulse accuracy between the Cellnovo and OmniPod devices, highlighting the importance of pumping mechanisms.
Area of Science:
- Endocrinology
- Biomedical Engineering
- Medical Devices
Background:
- Insulin patch pumps represent a novel approach to insulin delivery for diabetes management.
- Accurate insulin delivery is critical for maintaining glycemic control and preventing complications.
Purpose of the Study:
- To conduct a comparative analysis of the pulse accuracy between two distinct insulin patch pump systems.
- To evaluate the impact of different pumping mechanisms on device precision.
Main Methods:
- Micro-gravimetric evaluation was employed to assess the accuracy of insulin delivery from two patch pump devices: Cellnovo and OmniPod.
- Pulse accuracy was determined by analyzing both single and time-averaged insulin pulses for each device.
Main Results:
- A higher percentage of single pulses from the OmniPod fell outside accuracy thresholds (±5% to ±30%) compared to the Cellnovo device.
- For averaged pulse windows (10, 20, and 40 pulses), the Cellnovo demonstrated significantly better accuracy (mean values outside ±15%) than the OmniPod.
Conclusions:
- The study demonstrates variability in accuracy among insulin patch pump technologies.
- The specific pumping mechanism utilized by a patch pump significantly influences its accuracy and precision in insulin delivery.
More Related Videos
07:21Subcutaneous Angiotensin II Infusion using Osmotic Pumps Induces Aortic Aneurysms in Mice
Published on: September 28, 2015
10:19Neurodegeneration in an Animal Model of Chronic Amyloid-beta Oligomer Infusion Is Counteracted by Antibody Treatment Infused with Osmotic Pumps
Published on: August 14, 2016
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Improving Translational Accuracy
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
ATP Driven Pumps II: P-type Pumps
A typical P-type pump has three cytosolic domains: nucleotide-binding (N), phosphorylation (P), and activator (A) domains. These domains are connected to the membrane-spanning helices by short amino acid segments. ATP hydrolysis and covalent phosphoenzyme intermediate formation are crucial parts of the catalytic cycle. At the highly...